SIEMENS SITRANS P200 7MF1565 Pressure transmitter
The SITRANS P200 pressure transmitter measures the gauge and absolute pressure of liquids, gases and vapors.
- With ceramic measuring cell
- Gauge and absolute measuring ranges 1 to 60 bar (15 to 1000 psi)
- For general applications
Benefits
- High measuring accuracy
- Rugged stainless steel enclosure
- High overload withstand capability
- For aggressive and non-aggressive media
- For measuring the pressure of liquids, gases and vapors
- Compact design
Application
The SITRANS P200 pressure transmitter for gauge and absolute pressure is used in the following industrial areas:
- Mechanical engineering
- Shipbuilding
- Power engineering
- Chemical industry
- Water supply
Design
Device structure without explosion protection
The pressure transmitter consists of a piezoresistive measuring cell with a diaphragm installed in a stainless steel enclosure. It can be connected to the electricity supply using a connector to EN 175301-803-A (IP65), a round plug M12 (IP67), a cable (IP67) or a Quickon cable quick screw connection (IP67). The output signal is between 4 and 20 mA or 0 and 10 V.
Device structure with explosion protection
The pressure transmitter consists of a piezoresistive measuring cell with a diaphragm installed in a stainless steel enclosure. It can be used with a connector per EN 175301-803-A (IP65) or a round plug M12 (IP67) connected electrically. The output signal is between 4 and 20 mA.
Function
The pressure transmitter measures the gauge and absolute pressure of liquids, gases and vapors.
Mode of operation
SITRANS P200 pressure transmitters (7MF1565-…), functional diagram
The ceramic measuring cell has a thin-film resistance bridge to which the operating pressure p is transmitted through a ceramic diaphragm.
The voltage output from the measuring cell is converted by an amplifier into an output current of 4 to 20 mA or an output voltage of 0 to 10 V DC.
The output current and voltage are linearly proportional to the input pressure.
Product No. / Product Description | SITRANS P210 for gauge pressure Ordering data | ||||||
![]() | 7MF1565-…..-…. SITRANS P200 Transmitters for pressure and absolute pressure for general applications Non-linearity: 0,25 percent (typically) wetted parts material:ceramics, stainless steel plus sealing material; non-wetted parts material: stainless steel |
SIEMENS SITRANS P200 Technical specifications
Application | |
Gauge and absolute pressure measurement | Liquids, gases and vapors |
Mode of operation | |
Measuring principle | Piezo-resistive measuring cell (ceramic diaphragm) |
Measured variable | Gauge and absolute pressure |
Input | |
Measuring range | |
| |
| 1 … 60 bar (15 … 870 psi) |
| 15 … 1000 psi |
| |
| 0.6 … 16 bar a (10 … 232 psia) |
| 10 … 300 psia |
Output | |
Current signal | 4 … 20 mA |
| (UB – 10 V) / 0.02 A |
| 7 … 33 V DC (10 … 30 V for Ex) |
Voltage signal | 0 … 10 V DC |
| ≥ 10 kΩ |
| 12 … 33 V DC |
| < 7 mA at 10 kΩ |
Radiometric output | 10 … 90% |
| ≥ 10 kΩ |
| DC 5 V ± 10% |
| < 7 mA at 10 kΩ |
Characteristic curve | Linear rising |
Measuring accuracy | |
Error in measurement at limit setting including hysteresis and reproducibility |
|
Step response time T99 | < 5 ms |
Long-term stability | |
| 0.25 % of full-scale value/year |
Effect of ambient temperature | |
| 0.25 %/10 K of full-scale value |
| 0.005 %/V |
Rated conditions | |
Pressure transmitter with gasket made of: | |
| -15 … +125 °C (5 … 257 °F) |
| -35 … +100 °C (-31 … +212 °F) |
| -20 … +100 °C (-4 … +212 °F) |
| -40 … +145 °C (-40 … +293 °F), usable for drinking water |
Ambient temperature | -25 … +85 °C (-13 … +185 °F) |
Storage temperature | -50 … +100 °C (-58 … +212 °F) |
Degree of protection according to IEC 60529 |
|
Electromagnetic compatibility |
|
Design | |
Weight | Approx. 0.090 kg (0.198 lb) |
Process connections | See dimension drawings |
Electrical connections |
|
Wetted parts materials | |
| AI2O3 – 96 % |
| Stainless steel, mat. no. 1.4404 (SST 316 L) |
|
|
Non-wetted parts materials | |
| Stainless steel, mat. no. 1.4404 (SST 316 L) |
| Plastic |
| PVC |
Certificates and approvals | |
Classification according to pressure equipment directive (PED 2014/68/EU) | For gases of fluid group 1 and liquids of fluid group 1; complies with requirements of article 4, paragraph 3 (sound engineering practice) |
Lloyd’s Register of Shipping (LR)1) | 12/20010 |
Germanischer Lloyd (GL)1) | GL19740 11 HH00 |
American Bureau of Shipping (ABS)1) | ABS_11_HG 789392_PDA |
French marine classification society (BV)1) | BV 271007A0 BV |
Det Norske Veritas (DNV)1) | A 12553 |
Drinking water approval (ACS)1) | ACS 15 ACC NY 360 |
EAC1) | № TC RU C-DE.ГБ05.В.00732 OC НАНИО «ЦСВЭ» |
Underwriters Laboratories (UL)1) | |
| UL 20110217 – E34453 |
| IEC UL DK 21845 |
Explosion protection | |
Intrinsic safety “i” (only with current output) | Ex II 1/2 G Ex ia IIC T4 Ga/Gb Ex II 1/2 D Ex ia IIIC T125 °C Da/Db |
EC type-examination certificate | SEV 10 ATEX 0146 |
Connection to certified intrinsically-safe resistive circuits with maximum values | Ui ≤ 30 V DC; Ii = 100 mA; Pi ≤ 0.75 W |
Effective internal inductance and capacity for versions with plugs per EN 175301-803-A and M12 | Li = 0 nH; Ci = 0 nF |
1) For variants with output signal 0 … 5 V and radiometric output available soon.
Dimensional drawings
SITRANS P200, electrical connections, dimensions in mm (inch)
SITRANS P200, process connections, dimensions in mm (inch)
Circuit diagrams
Connection with current output and connector per EN 175301
Connection with current output and connector M12x1
Connection with current output and cable
Connection with current output and Quickon cable quick screw connection
Connection with voltage output, ratiometric output and plug according to EN 175301
Connection with voltage output, ratiometric output and M12x1 plug
Connection with voltage output, ratiometric output and cable
Connection with voltage output, ratiometric output and Quickon fast cable termination
Version with explosion protection: 4 … 20 mA
The grounding connection is conductively bonded to the transmitter enclosure.
Connection with current output and connector per EN 175301 (Ex)
Connection with current output and connector M12x1 (Ex)