ACCURATE pH MEASUREMENT
Electrode selection
In order to make precise pH measurements, there are a number of pre-requisites. Special attention should be paid to the choice of the sensors to ensure compatibility with the chemical
composition of the process fluid. The speed of response required, the solids content and the flow rate of the fluid are also contributory factors. Other Yokogawa specification sheets cover
the choice of sensors. Any Yokogawa sales office can provide expert advice.
Converter design
The system should be designed to minimize the effect of external influences. The EXA series features dual high impedance input amplifiers in a differential configuration with liquid earth. This
system is used to ensure compatibility with most available sensor combinations. It also provides excellent noise rejection, minimizing the stray signals that can affect industrial pH
measurements. Earth loop currents in damp and damaged cabling are eliminated by the equipotential screening. EXA instruments offer a simple and effective process temperature
compensation in addition to the usual compensation to the Nernst equation.
Maintenance
It is important that the system be well maintained. The electrodes must be properly cleaned and regularly calibrated. Yokogawa online cleaning systems may be used where there is significant
fouling of the sensors. Other influences from the electrode holders can be less obvious, but important none the less. Well designed fittings make it easy to provide the routine maintenance
needed for best accuracy. Flow through, retractable and immersion assemblies are available.
The selection of the buffer solutions is important to ensure the best possible calibration. Yokogawa strongly recommends the use of NIST (NBS) buffers. These are chosen as primary buffer
standards because of the maximum buffer capacity exhibited by these solutions. Commercial buffers (adjusted with NaOH) have the limited advantage of an integer value, with the disadvantages
of loss of buffer capacity and greater temperature dependence.
Yokogawa sets the pH / temperature relationships for NIST buffers in the software of the EXA pH converters. This combination provides the best calibration possible.
Accessories
Buffer powder packs to NIST (NBS) standard. Each box contains
five (5) packets, each sufficient to make 200 ml of calibration
solution when dissolved in distilled water.
SIGNAL PROCESSING (pH/ORP/rH)
With the correct sensor configuration it is possible to measure pH and ORP or rH (ORP is the oxidation reduction potential including the pH-component, rH compensates for the effect of pH).
Combinations of two parameters can be displayed and transmitted separately on mA2.
Communications and Logbook
The EXA PH402 is equipped with RS485 communication ability. Communication lines are isolated from the input and output signals. Communication speed is selectable from 1200, 2400,
4800, 9600 baud The format is selectable for even, odd, and no parity. The PH402 can be accessed over this 2-way communication link.
In combination with 2-way communication, there is a logbook feature. This is a software record of events, calibrations, alarms and the like. A total of approximately 100 events can be stored.
New additions are made by overwriting the oldest entry, once 100 entries have been exceeded. Access is via the RS485 link, and Yokogawa’s, PC402 software package. Important diagnostic
information can also be read on the LCD display.