PH450G-A(D)-A
The PH450 is supplied with terminals suitable for the connection of finished wires in the size range of 0.13 to 2.5 sq.mm. ( 26 to 14 AWG).
The cable glands supplied will form a tight seal on cables with an outside diameter of 6 to 12 mm (0.24 to 0.47 inches). Unused cable entry holes must be sealed with cable glands including the close up plugs supplied.
PH450G-A(D)-U
The PH450 is supplied with terminals suitable for the connection of finished wires in the size range of 14- 26 AWG. The cable entry holes are sealed with FM certified plugs. Prior to cable entry the plugs can be removed with allen key size 3/8” The cable conduit fittings can be mounted in the holes of the housing as required. The cable glands supplied with the unit will give a tight seal on cables with outside diameter of 0.24 to 0.47 inches.
GENERAL SPECIFICATIONS OF EXAXT PH450G
A) Inputs specs : Dual high impedance input (≥ 1012Ω)
(under referance conditions)
B) Input ranges
pH : -2 to 16 pH
ORP : -1500 to 1500 mV
rH : 0 to 100 rH
Temperature
– Pt1000 : -30 to 140ºC
– Pt100 : -30 to 140ºC
– 350Ω (DKK) : -30 to 140ºC
– 5k1 : -30 to 140ºC
– 6k8 : -30 to 140ºC
– PTC10k : -30 to 140ºC
– NTC 8k55 : -10 to 120ºC
– 3kBalco : -30 to 140ºC
C) Accuracy
pH input : ≤ 0.01 pH
ORP input : ≤ 1 mV
Temperature : ≤ 0.3 ºC (≤ 0.4 ºC for Pt100)
Step response : < 4 sec for 90% (pH 7 – pH 4) Note on performance specifications The following tolerance is added to above performance. mA output tolerance: ±0.02 mA of “4-20 mA“ D) Transmission signals General : Two isolated outputs of 4-20 mA. DC with common negative. Maximum load 600Ω. Bi-directional HART® digital communication, superimposed on mA1 (4-20 mA) signal Output functions : Linear or Non linear (21-step table) output for pH, temperature, ORP or rH Control function : PID control Burnout function : Burn up (21.0 mA) or burn down (3.6 mA) to signal failure acc. NAMUR NE43 Adjustable damping. Expire time Hold : The mA-outputs are frozen to the last/fixed value during calibration/ commissioning E) Contact outputs General : Four SPDT relay contacts with display indicators Switch capacity : Maximum values 100 VA, 250 VAC, 5 Amps.(*) Maximum values 50 Watts, 250 VDC, 5 Amps.(*) Status : High/Low process alarms, selected from pH, ORP, rH and temperature Configurable delay time and hysteresis Failure annunciation Control function : On/Off, PID duty cycle or pulsed frequency control Wash : Contact can be used to start manual- or interval time wash cycles Hold : Contact can be used to signal the Hold situation. Fail : Contact S4 is programmed as fail-safe contact (*) Note: When contact output current is more than 4 Amps. ambient temperature should be less than 40ºC. F) Contact input : Remote wash cycle start. G) Temperature compensation Function : Automatic or manual Compensation to Nernst equation Process compensation by configurable temperature coefficient, NEN6411 for water or strong acids/bases or programmable matrix H) Calibration : Semi-automatic 1 or 2 point calibration using pre-configured NIST, US, DIN buffer tables 4, 7 & 9, or with user defined buffer tables, with automatic stability check Manual adjustment to grab sample I) Logbook : Software record of important events and diagnostic data readily available in the display or through HART®. J) Display : Graphical Quarter VGA (320 x 240 pixels) LCD with LED backlight and touchscreen. Plain language messages in English, German, French, Spanish, Italian, Swedish ,Portuguese and Japanese. K) Shipping details Package size : 290 x 300 x 290 mm (L x W x D) (11.5 x 11.8 x 11.5 inch) Package weight : app. 2.5 kg (5.5 lbs) Converter weight : app. 1.5 kg L) Housing : Cast aluminum housing with chemically resistant coating; Polycarbonate cover with Polycarbonate flexible window : Protection IP66 / NEMA 4X / CSA Type 3S Colour : Silver grey PH450G-A(D)-A : IP66 cable glands are supplied with the unit PH450G-A(D)-U : NEMA 4X close up plugs are mounted in the unused cable entry holes and can be replaced by conduit fittings as required Pipe, Panel or Wall mounting using optional hardware Optional conduit adapter G1/2, 1/2NPT or M20 female M) Power supply PH450G-A : Ratings; 100-240 V AC Acceptable range; 90 to 264 V AC Ratings; 50/60 Hz Acceptable range; 50 Hz ±5%, 60 Hz ±5% Power Consumption; 15 VA PH450G-D : Ratings; 12-24 V DC Acceptable range; 10.8 to 26.4 V DC Power Consumption; 10 W N) Safety and EMC conforming standards , Safety : EN 61010-1 CSA C22.2 No.61010-1 UL 61010-1 FM3611 Class I, Div.2, Group ABCD,T6 for Ta -20 to 55ºC EMC : EN 61326-1 Class A, Table 2 (For use in industrial locations) EN 61326-2-3 EN 61000-3-2 Class A EN 61000-3-3 Korea Electromagnetic Conformity Standard Installation altitude: 2000 m or less Category based on IEC 61010: II (Note) Pollution degree based on IEC 61010: 2 (Note) Note: Installation category, called over-voltage category, specifies impulse withstand voltage. Category II is for electrical equipment. Pollution degree indicates the degree of existence of solid, liquid, gas or other inclusions which may reduce dielectric strength. Degree 2 is the normal indoor environment.